Water Quality Monitoring

Data Collection Methods - Turbidity

Level 1-2
Level 3-4

Procedure

This test is made by comparing the turbidity of a measured amount of the sample with an identical amount of turbidity-free water containing a measured amount of standardized turbidity reagent. The readings are made by looking down through the column of liquid at a black dot. If turbidity is present, it will interfere with the passage of light through the column of liquird. Small amounts of turbidity will cause a "blurring" of the black dot in the bottom of the tube. Large amounts of turbidity may provide sufficient "cloudiness" so that it is not possible to see the black dot when looking down through the column. Any color that may be present in the sample should be disregarded. This determination is concerned only with the haziness or cloudy nature of the sample.

  1. Fill one Turbidity Column (0835) to the 50 ml. line with the sample water. If the black dot on the bottom of the tube is not visible when looking down through the column of liquid, pour out a sufficient amount of the test sample so that the tube is filled to the 25 ml. line.

  2. Fill the second Turbidity Column (0835) with an amount of turbidity-free water that is equal to the amount of sample being measured. Distilled water is preferred; however, clear tap water may be used. This is the "clear water" tube.

  3. Place the two tubes side by side and note the difference in clarity. If the black dot is equally clear in both tubes, the turbidity is zero. If the black dot in the sample tube is less clear, proceed to Step 4.

  4. Shake the Standard Turbidity Reagent (7520) vigorously. Add 0.5 ml. to the "clear water" tube. Use the sitting rod (1114) to stir the contents of both tubes to equally distribute the turbid particles. Check for the amount of turbidity by looking down through the solution at the black dot. If the turbidity of the sample water is greater than that of the "clear water", continue to add Standard Turbidity Reagent in 0.5 ml. increments to the "clear water" tube, mixing after each addition until the turbidity equals that of the sample. Record total amount of Turbidity Reagent added.

  5. Each 0.5 ml. addition to the 50 ml. size sample is equal to 5 Jackson Turbidity Units (JTU's). If a 25 ml. sample size is used, each 0.5 ml. addition of the Standard Turbidty Reagent is equal to 10 JTU's. (See the table below.) Rinse both tubes carefully after each determination.

Turbidity Test Results

Number of Measured Additions Amount in ml. 50 ml. Sample 25 ml. Sample
1 0.5 5 JTU 10 JTU
2 1.0 10 JTU 20 JTU
3 1.5 15 JTU 30 JTU
4 2.0 20 JTU 40 JTU
5 2.5 25 JTU 50 JTU
6 3.0 30 JTU 60 JTU
7 3.5 35 JTU 70 JTU
8 4.0 40 JTU 80 JTU
9 4.5 45 JTU 90 JTU
10 5.0 50 JTU 100 JTU
15 7.5 75 JTU 150 JTU
20 10.0 100 JTU 200 JTU

1/96

LaMotte Company, P.O. Box 329, Chestertown, MD 21620, USA
1-800-344-3100
email: ese@lamotte.com